Abstract
Study Design : comparative cross-sectional study
Purpose : To compare the Anterior Segment (AS) parameters obtained by Pentacam and IOL master 700 in Indian population.
Patients and Methods: This study included 80 eyes of 72 Indian individuals. ACD, K-readings and WTW values were measured with IOL master 700 and Pentacam Oculus and compared. All investigations and examinations were done at a tertiary care centre from January 2023 to June 2023.
Results: IOL Master 700 gave 43.56± 1.72 & 45.03 ± 1.59 and Pentacam 43.28± 1.73 & 44.72± 1.58 for K1 & K2 respectively. Low significant difference between both devices in values of ACD (mean difference ± SD – 0.03± 0.10). Pentacam measures longer WTW value (mean 12.10± 0.45) than IOL master (mean 12.00±0.41).
Conclusion: IOL master 700 gives slightly higher K-reading than Pentacam, which is statistically significant. Both devices give nearly similar results for ACD & WTW.